Instrument operating instructions and matters needing attention
⑴The instrument must be reliable grounding, power entrance lead into AC220V power.
⑵Open tank cover, and take out the test cell, we should shut the tank cover, when heating and testing the dielectric loss.
⑶The cover is protective and when you open it, will interrupt the heating and high pressure.
⑷During the test, there is high pressure and high temperature. Do not contact the test cell, cable and socket when the power is electrify and testing.
⑸Take care in order to avoid sprinkling the oil into the slot of test cell and the top panel when injecting and discharging oil.
⑹If it appears dead stop faults phenomenon while testing, turn the system back on, and restart the instrument.
Use the appropriate power supply cables.
Only use the product-specific power supply line in accordance with the product specification. Be right to disconnect and connect.
The instrument must be adequately grounded.
In addition to the power line grounding conductor grounding, product shell’s grounding columns must be grounded
In order to prevent electric shock, the grounding conductor must be connected with the ground.
Before connecting with the input or output terminals of the product, make sure that the product is properly grounded.
Note that rating value of all terminals.
To prevent fire or electric shock hazard, please note that all rating values and markings in the product.
Before connecting this product, please read the product manual in order to further understand the information about rating values.
Use the appropriate protective tube.
Use only the protective tube in accordance with provisions of this product type and rating.
Do not operate when there is any suspicion of failure. If you suspect this product is damaged, please contact our repair staff to check, and do not continue the boot operation.
Align the locating slot when the internal electrode into the external electrode.
Do not operate in humid environments.
Do not operate in explosive environments.
Keep the surface clean and dry.
Special notice: This instrument has high voltage output, improper use can compromise personal safety, prior to the use of this instrument, be sure to read the instruction manual carefully!

Technical Data
Voltage supply: AC 220V±10%
Power frequency:50Hz/60Hz ±1%
Measuring range:
Electric capacity 5pF~200pF
Relative Permittivity 1.000~30.000
Dielectric loss factor 0.00001~100
DC Resistivity 2.5MΩm~20TΩm
Measuring accuracy:
Electric capacity ± (1% of measured value+0.5pF)
Relative Permittivity ±1% of measured value
Dielectric dissipation factor ± (1% of measured value+0.0001)
DC Resistivity ±10% of measured value
Resolution:
Electric capacity 0.01pF
Relative capacitivity 0.001
Dielectric dissipation factor 0.00001
Measured temperature range:0~120℃
Temperature measurement error:±0.5℃
Test Voltage AC(RMS):500~2000V Continuously adjustable, frequency 50Hz
Test Voltage DC:300~500V Continuously adjustable
Power consumption:100W
External dimensions:500×360×420
Total weight:22Kg

Huazheng Electric Manufacturing Co., Ltd is founded in 2008,the supplier of the State Grid in China.It means we are the supplier of the Chinese Government.And our products spread throughout the whole China.
Our main products : Transformer Oil Tester, Cable Fault Locator, High Voltage Generator,transformer tester, relay protection tester,insulation resistance tester,Power Quality Analyzer and so on.
Close to Tianjin port, the transportation is convenient. our customers come from More than 50 countries, like America, England, Bulgaria, Brazil, Chile, Malysia, Vietnam and so on.
What's more, gurantee 1 year,life long maintance .Products through the CE, ISO, IEC certification, quality and reliable, factory price.
Warmly welcome all partners to join us for mutual benefits cooperation.








所有评论仅代表网友意见,与本站立场无关。