ASTM D93 Oil Flashpoint Analyzer Closed Cup Flash Point Measuring Instrument

I.Function Overview
Flash point measuring instrument, Used to determine the closed flash point value of petroleum products. Adopt advanced foreign technology, convenient and fast, open, fuzzy control integrated software, modular structure, in line with national standards, the United States and other standards. It is an ideal substitute for imported instruments.
The equipment under the conditions stipulated in the national standard, put the sample into test cup, affect the heating experiment of oil cup, with the oil vapor ambient air and gas with the lowest temperature will ignite a flame as a flash point. The digital signal processor controls the heater according to the temperature change, so that the test oil temperature rises at a certain rate. Automatic control of ignition and flash detection; When the flash is measured, the system stops data collection, displays the flash temperature, prints the recording results, and stops heating.
III.Technical Parameters
Temperature Measurement
Temperature measuring range: -49.9°C~400.0°C
Repeatability: 0.029X(x=twice successive test average value)
Resolution: 0.1°C
Precision: 0.5%
Temperature Sensor: Platinum Resistor(PT100)
Flash fire detection: K Thermocouple
IV.Instrument Structure And Installation
1)Instrument Installation
A.Unpack the instrument and make sure that the equipment is in good condition.
B.Check instrument model and all accessories according to the packing list.
C.Please debug the instrument after inspection step 1 & 2.
2)Starting-up Preparation
A.Connect the instrument to the AC 220V power supply;
Note: Always use a power outlet with a reliable ground to power the instrument;
B.Clean the oil cup with petroleum ether, pour sample liquid into the oil cup to the scale mark, and put into the heater.
C.Correctly install the oil lamp and platinum alloy electric ignition wire ignition head.