HZJF90V1 PD Measurment Partial Discharge Meter
This advanced instrument utilizes high-frequency detection technology to measure partial discharge (PD) in electrical equipment. It integrates cutting-edge technologies including:
Sensing: Ultrasonic, Radio Frequency (RF), Transient Earth Wave (TEW), High Frequency (HF), and Very High Frequency (VHF) detection.
Processing: High-speed signal acquisition, analog electronics, transient waveform sampling, and advanced digital signal processing.
Platform: Linux-based operating system for robust performance and future IoT integration.
Key Capabilities:
Generates 2D/3D PD maps, amplitude-phase spectra (PRPD/PRPS), and quantified discharge values.
Enables detailed static or dynamic analysis of single/multi-cycle PD pulse waveforms for accurate equipment condition assessment.
Ideal for live detection and daily insulation inspection of high-voltage cables, transformers, and other critical assets.
Advantages:
Highly Sensitive & Adaptable: Effectively identifies insulation defects.
Portable Design: Compact, lightweight, and easy to use on-site.
Robust Performance: Features superior anti-interference and measurement accuracy via RF/high-frequency/weak-signal processing and high-precision AD conversion.
Main Features & Indicators:
Rapid On-Site PD Detection: Specifically optimized for high-voltage cables; portable and user-friendly.
Multi-Method Detection: Identifies PD signals via Ultrasonic, TEW, HF, and VHF.
Flexible Sensing: Built-in non-contact air-coupled sensor; supports external surface-coupled ultrasonic sensors.
Comprehensive Analysis: Multiple analysis windows (Phase-Linear, PRPD, PRPS).
Data Management: Store, view, and delete test data; supports waveform recording.
Intelligent Alerting: Configurable alarm thresholds with color-coded amplitude indicators for clear issue identification.
Extended Operation: Large-capacity battery provides 6-8 hours continuous use.
Future-Ready Platform: Linux OS facilitates easy feature expansion and IoT integration.
Main Technical Parameters
1. Non-Contact Ultrasonic Sensor (AA)
Central Frequency: 40 kHz ±1 kHz
Measurement Range: -7 to 68 dBµV
Resolution: 0.1 dB
Form Factor: Interchangeable Bendable/Focusing Design
2. Contact Ultrasonic Sensor (AE)
Detection Band:
SF6 Equipment: 20–80 kHz
Oil-Filled Equipment: 80–200 kHz
Dynamic Range: 60 dB
Resolution: 0.1 dB
3. Transient Earth Voltage Sensor (TEV)
Detection Band: 3–100 MHz
Measurement Range: 0–60 dBmV
Resolution: 0.1 dB
4. Ultra-High Frequency Sensor (UHF)
Detection Band: 300–1500 MHz
Measurement Range: -70 to -10 dBm
Resolution: 0.1 dBm
Effective Height: ≥10 mm
5. High-Frequency Sensor (HF)
Detection Band: 0.3–30 MHz
Measurement Range: 0–70 dB
Gain Range: -30 to 20 dB
Resolution: 0.1 dB
Impedance:
Input: 50 Ω
Transmission: ≥10 mV/mA
6. Device Parameters
Operating Environment:
Temperature: -20°C to 50°C
Humidity: 0–85% RH (non-condensing)
Display: 5.6" TFT LCD (480 × 640 px)
Dimensions: 230 × 170 × 48 mm
Weight: 1.35 kg
Battery Life: ≥6 hours continuous operation