HZ6DPlus Automatic Infrared Spectrophotometer
This automatic oil measuring instrument is composed of automatic operation software, infrared spectroscopy system and magnetic stirring extraction system.
The instrument uses the extraction solvent (CCl4) according to the requirements of the national standard to extract the oil in the water body with the magnetic side three-dimensional side magnetic stirring technology according to a certain extraction ratio. Then the infrared spectroscopy system analyzes and detects the oil content in the extraction solution. If a special magnesium silicate filter device is installed, the content of petroleum and animal and vegetable oils can also be measured. After the analysis and measurement, the waste liquid is discharged to clean the entire pipeline to ensure that the next water sample will not be polluted repeatedly.
The above process is automated through the automatic operation of intelligent software, and the detection of oil in the sample water can be completed without the operator touching tetrachloroethylene, that is, automatic sample injection, automatic extraction, automatic water removal and impurity removal, automatic measurement, automatic drainage, Automatically cleans and stores data.
Main Data
1. The detection limit of the instrument is 0.02mg/L (measure 11 times of blank and calculate 3 times SD)
2. Wave number range 3400cm-1~2400cm-1 (ie 2941nm~4167nm)
3. Absorbance range 0.0000~2.0000AU (ie transmittance 100~1%T)
4. The minimum detection concentration is 0.002mg/L (enrichment extraction method)
5. Maximum measurement concentration 10000mg/L (dilution measurement)
6. Instrument measuring range (0.05~180) mg/L
7. Repeatability RSD ≤1% (20-100mg/L oil standard sample was measured 11 times)
8. Measurement accuracy error <±5 %
9. Linear correlation coefficient R>0.999
10. Water sample volume 5ml—600ml
11. The extraction amount of tetrachloroethylene is an integer multiple of 10 or 20ML
12. Automatic detection of a single sample 7-18min (the more the sample volume, the longer the extraction time)
13. Wavenumber accuracy ±1cm-1
14. Host net weight 25kg
15. Use power supply (220±22)V, (50±1)Hz, 50VA
16. Use temperature and humidity temperature range 1℃-40℃; humidity≤80%
The standard basis for the use of this product:
Application field | Standard |
Determination of environmental monitoring stations, domestic sewage and industrial wastewater | GB8978—1996 Integrated Wastewater Discharge Standard |
Fume monitoring facility of catering industry | GB18483—2001 Food and Beverage Industry Fume Emission Standard |
Municipal drainage testing station or sewage treatment plant | GB18918—2002 Pollutant Discharge Standards for Urban Sewage Treatment Plants |
Town water company or water company | GB3838—2002 Environmental Quality Standard for Surface Water; 2000 Urban Drinking Water Hygiene Standard |
Petroleum and petrochemical enterprises | GB3551—83 Discharge Standard for Water Pollutants in Petroleum Refining Industry |
Railway Environmental Monitoring Station | GB5469—85 Discharge Standard for Wastewater from Railway Freight Car Washing |
Thermal Power Plant Environmental Monitoring Station | GB8978—1996 Comprehensive sewage discharge standards and other standards |
Wastewater monitoring in the smelting industry | GB13456—92 Discharge Standard for Water Pollutants in Iron and Steel Industry |
Marine fishery water quality monitoring | GB3097 Marine water quality standard and GB11607 fishery water quality standard |
Hydrological Environment Monitoring Station | GB3838—2002 Surface Water Environmental Quality Standard |
Agricultural Product Quality Supervision and Inspection Center, Agricultural Forestry Environmental Monitoring Station | GB4284—84 Standard for Pollutant Control in Agricultural Sludge |