DAC-IR-2C Silicon Steel Sheet Iron Loss Tester
DAC-IR-2C is used for iron loss test of electrical iron plate (silicon steel sheet). It is easy to carry, easy to test and good repeatability. As long as the probe is placed on the test piece, the iron loss value of the test piece can be measured in the form of W/kg and displayed digitally.
(1) Main applications:
Incoming and outgoing test of silicon steel sheet
Quality control during calcination
Quality management of silicon steel sheet, etc
(2) Equipment features:
Non-destructive measurement method, and can directly read its iron loss value
Iron loss value is displayed in W/kg
The touch digital setting switch is adopted for easy and clear operation
When the probe leaves the sample, it automatically stops the test to avoid long-time operation
Use standard iron plate to facilitate the calibration of probe
Automatic control of magnetic flux density
Strong anti-power noise function, no need to use AC stable power supply
General Description
Measurable Watt Loss: 0.10-19.99 W/kg (4 digits)
Magnetic Flux Density: 1.5T and 1.7T (Tesla) for Non-Oriented(S) and Oriented Grain(G) Thickness Selection: By Setting Button (3 digits)
Width Selection: By Setting Button (3 digits)
Possible to measure: 0.10-0.90mm (Thickness); Wider than 5mm(Width) and 20mm(Length) Test Object: Electrical Steel Sheets for Non-Oriented or Oriented Grain.
Measuring Accuracy: Error less than±5%.
Measuring Frequency: 50Hz (synchronized to power supply frequency) and 60Hz(option) Input Power Supply: AC100V and 220V selected according to request by customer.
Size: 220(W)×140(H)×350(D)mm
Net Weight: About 4 KGS