250V To 5kV Digital Insulation Resistance Tester Meter Meger Price





I.Features
MODLE HZJY-5K-I are high voltage insulation resistance testers with 5-range(HZJY-5K-I)for measuring insulation resistance.
● Designed to meet the following safety standards:
IEC 61010-1,-2-30 (CAT III 600V/CAT IV 300V Pollution degree 2)
IEC 61010-031 (Requirements for hand-held probes)
● With auto-discharge function
When an insulation resistance like a capacitive load is measured, electric charges stored in capacitive circuits are automatically discharged after the measurement. Discharge can be checked on the voltage monitor.
● Backlight function to facilitate working at dimly illuminated location or at nighttime work.
● Bar graph display
● Audible and visible LIVE circuit warning
● With Auto-power off function
The instrument automatically will turn itself off if there is no function change or button press for 10 min to conserve battery power.
● Auto-measurement and display of PI (Polarization Index), DAR(Dielectric Absorption Ratio)
II.Specification
● Measuring range and accuracy
(Temperature, humidity;23±5℃,45-75%RH)
【Insulation resistance tester】 : MODLE HZJY-5K-I
Rated voltage | 250V | 500V | 1000V | 2500V | 5000V |
MeasuringRange | 0.1~100.0MΩ | 0.1~99.9MΩ 80~1000MΩ | 0.1~99.9MΩ 80~999MΩ 0.80~2.00GΩ | 0.1~99.9MΩ 80~999MΩ 0.80~9.99GΩ 8.0~100.0GΩ | 0.1~99.9MΩ 80~999MΩ 0.80~9.99GΩ 8.0~99.9GΩ 80~20000GΩ |
Display range | 0.1~100.0 MΩ | 0.1~1000 MΩ | 0.1M~2.00 GΩ | 0.1M~100.0 GΩ | 0.1M~2000 GΩ |
Open circuit Voltage | DC250V +10% ,-10% | DC 500V + 20%,-10% | DC 1000V + 20%,-0% | DC 2500V +20%,-0% | DC5000V +20%,-0% |
Rated Current |
|
| 1MΩ 负荷时 1mA~1.2mA | 2.5MΩ 负荷时 1mA~1.2mA | 5MΩ 负荷时 1mA~1.2mA |
Short-circuit Current | 5mA±0.5mA | ||||
Accuracy | ±5%±3dgt | ±5%±3dgt±20% | |||
Voltage monitor for insulation resistance range
MODLE HZJY-5K-I:30 – 6000V (resolution 10 V): ±10%rdg±20V
This monitor is used to check whether electric charges stored no the equipment under test are discharged or not. The voltage value measured and displayed on the LCD is a reference value. Please note that the indicated value, when an external AC voltage is applied to the instrument, is not a correct value.
【Voltage measurement range】
| DC voltage | AC voltage |
Measuring range | ±30-±600V | 30-600V(50/60Hz) |
Resolution | 1V | |
Accuracy | ±2%rdg±3dgt | |
● Display: Liquid crystal display insulation resistance range:(Max.1200 counts) VAC/DC range: (Max.630counts) Bar graph: Max.36points DAR/PI value:Max.9.99 Time:Max.99.59
● Low battery warning: Battery symbol (in 4 levels)
● Overrange indication: “OL" mark appears on insulation resistance range.
● Auto-ranging: Range shifts to upper range:1000 counts
Range shifts to lower range:80 counts
(Only on the insulation resistance range)
● Auto-power-off: The instrument will be powered off if there is no switch or button operation for 10 min.(This function does not work during a measurement.)
● Altitude: 2000m or less
● Temperature & humidity range(guaranteed accuracy):
23℃±5℃/Relative humidity 85% or less(no condensation)
● Operating temperature & humidity range:
0℃ to 40 ℃/Relative humidity 85% or less (no condensation)
● Storage temperature&humidity range:
-20℃ to ± 60%/ Relative humidity75% or less(no condensation)
● Overload protection: Insulation resistance range: AC 1200V/10sec. Voltage range:AC720V/10sec.
● Withstand voltage: AC5160V(50/60Hz)5sec.(Between electrical circuit and enclosure)
● Insulation resistance: 1000MΩ or more/DC1000V(Between electrical circuit and enclosure)
● Dimension: 177(L) ×226(W) ×100(D)mm
● Weight: MODLE HZJY-5K-I:approx. 1.9kg(battery included)
● Power source: DC12V: LR14(Alkaline battery size C)x8pcs
● Current consumption (representative values at 12V of supply voltage)
Range | 250V | 500V | 1000V | 2500V | 5000V | VAC/DC | ||||
Output at Short-circuit | 600mA | 110mA | ||||||||
When rated current is outputted | 650mA/0.25MΩ | 800mA/0.5MΩ | 1000mA/1MΩ | 1350mA/2.5MΩ | 1350mA/5MΩ | |||||
Output at open circuit | 40mA | 40mA | 50mA | 80mA | 150mA | |||||
Stand-by | 25mA | |||||||||
Backlight on | Increased by 40mA | |||||||||
CE Ceritification

● Measurement time: HZJY-5K-I:approx. for 25hours *under a load of 100MΩ
On the Insulation resistance 5000V range.
Accessories: Line probe: MODEL 7165 (Including MODEL8255
Straight type prod with molded parts)
Earth cord: MODEL 7264Guard
cord: MODEL 7265 LR14(Alkaline battery size C)x 8pcs instruction manual
Hard Case: MODEL 9179
Pickel Type Prod: MODEL 8019
Straight Type Prod: MODEL8254
● Optional accessories: Adaptor for recorder:MODEL8302 Line probe with
alligator clip:MODEL7168A
Long Line probe with alligator clip
(15m)