HZWS-C3 Karl Fischer Coulomb Titration Oil Water Content Tester
I.Overview
Karl Fischer Coulomb titration is the most reliable method to detect trace moisture for different substances, which was utilized successfully by trace moisture titrator. What’s important, this instrument is able to work more reliably and conveniently by using the most advanced control circuit and 32 bit embedded microcontroller processor as the master control core and utilizing embedded mini operating system. Character with fast analyzing speed, simple operation, high precision and strong automatism, the trace moisture titrator has been widely used in the industry of petroleum, chemical, power, railway, pesticide, medicine and environmental protection.
Instrument features:
●The 32-bit embedded microprocessor is used as the main control core to embed the mini operating system.
●0-400ma automatic detection, high precision, fast measurement speed, stable and reliable.
●The motherboard adopts high-quality SMT components with high integration, simple operation and long service life
●Contains 4 calculation formulas, and the measurement results are automatically converted to the required units and can be converted to each other.
●Color touch screen, full numeric keyboard, more concise operation, convenient and quick data calculation.
●With graph curve analysis to help analyze water content trend.
●Switching between Chinese and English at any time.
II.Technical parameters
Titration method:Coulometric titration(Coulomb analysis)
Display:Color LCD touch screen
Electrolytic current control:0~400mA automatic control
Measurement scope:1ug~200mg
Sensitive valve:0.1µg H2O
Accuracy:10µg~1000µg±3µg
More than 1mg but not greater than 0.3%
Printer:Miniature thermal printer
Power supply:220V±10%、50Hz
Power:< 40W
Ambient temperature:5~40℃
Ambient humidity:≤ 85%
External dimension:380×310×180
Weight:About 6kg
The determination of moisture in the sample
Before determining the operation, the following points should be determined whether they are correct:
1. Be sure to enter the measurement interface. 2. Is the voltage at the lower left corner stable? 3. Mix at the right speed.
1. Determination of moisture in liquid samples
(1) First, the 1ml injector with a needle (other capacity syringes can be selected according to the different samples on the side) is rinsed with the tested sample for 2-3 times, and then a certain amount of samples is taken to prepare for the injection.
(2). Press the Start button to clear the display. Tip: Please inject the sample to be tested for measurement
(3). The sample through into the cock in the anode chamber into the electrolyte, the attention should make needlepoint inserted into the electrolyte, and avoid contact with the wall or electrode, the titration will automatically start after injection, the determination of the end, the buzzer sounded, state prompt normal, digital display shows is the sample water content, unit of ug. Water content is shown to the right of water content .
Note 1: In the process of measurement, due to the error of operation, such as late clicking the back button, it will return to the starting interface and the measurement will be terminated, so the correct data cannot be obtained. In this case, it is necessary to re-enter the measurement interface until the state becomes normal and the equilibrium voltage is stable.