CS2671TX-1 AC DC 10kV 100mA Withstand Voltage Hipot Tester

1、Power switch
When the power switch is pressed, the tester is on, while pop-up it,the tester is off;
2、START key
In the reset status, press this key, the tester starts to perform test
3、STOP key
During the testing, it is used to interrupt the test. When DUT test failed, the failed indicator (10) is on, and at this time press the STOP key, the tester will stop alarming and enter into the next waiting test.
4、Remote interface
Use the remote probe to start up or reset of test remotely.
5、Voltage rotary knob
In the test of DUT, rotate this knob to regulate the output voltage (turn knob clockwise: voltage rising, while counterclockwise: voltage falling)
Confirm the rotary knob is at the position of “0" before power is on.
6、Loop terminal
It is the input terminal of measured current. During the test, the terminal leads should not be fallen off. If the test lead is fallen off, the DUT will be with high voltage and the electric shock maybe occurred.
7、High voltage output terminal (H.V.)
Press ‘test’ key, the terminal can generate high voltage. Do not touch the high voltage port, high voltage leads, high voltage probe and DUT.
8、Power supply safety detecting indicator light “O K"
Plug in, judge whether N、L、G are properly connected or not. The power switch is off, if the N、L、G is connected rightly, the “OK" indicator will light. If the N、L、G is wrongly connected, only “O" or “K" or neither of them is lighted. Please check the power supply. When power on, the indicator lights out.
9、High voltage output indicator
The indicator is not controlled by the main circuit. If the output voltage is more than 50V, the indicator will light up. If it is lighted on the reset status, the H.V. terminal maybe existed high voltage, please do not touch any part of the circuit. When the tester outputs high voltage, the indicator lights up all the time, which indicates it is dangerous during high voltage outputting.
10、Fail indicator
When the test is failed, the indicator will light.
11、Test indicator
When the start key is pressed, the indicator will light.
12、Voltage display
13、Current display
14、Time display
Time is range from 0.0s to 999s. When the test time is less than 100s, the resolution is 0.1s; while the time is greater than or equal to 100s, the resolution is 1s. If the time is set to 0.0s, the test time will add counts; when the time is not set to 0, the time will be down counting.
15、UP key
When do the test time set, press the up key, the test time set will be up.
16、DOWN key
When do the test time set, press the down key, the test time set will be down.
17、Current preset adjustment potentiometer
After pressing the numbered (18) button "test / Pre-set", the tester will enter into the current preset state, the current display will show the preset current value, clockwise to increase the preset current; counterclockwise to decrease the preset current.
18、2mA/20mA switch gear
Press the key is for 20mA gear, pop-up for 2mA gear.
19、Test /pre-set key
This key is pressed for the pre-set current state, while pop-up for test status.
20、100mA switch gear
Press the key is for 100mA gear, pop-up for 2mA/20mA gear.
21、5kV/10kV switch gear
Press the key is for 10kV gear, pop-up for 5kV gear.
| MODEL | Parameters |
| CS2671TX | ACW: 10kV, 100mA |
| CS2671TX-1 | ACW: 10kV, 100mA、DCW: 10kV, 20mA |
| A C W | Output voltage | (0.00~10.00)kV | |
| Max. output power | 1000VA | ||
| Max. current | 100mA | ||
| Current range | 2mA、20mA、100mA | ||
| Output waveform | sine wave | ||
| Waveform distortion | ≤5% | ||
| Test time | 0.0s~999s 0.0= continue | ||
| D C W | Output voltage | (0.00~10.00)kV | |
| Max. output power | 200VA | ||
| Max. current | 20mA | ||
| Current range | 2mA、20mA | ||
| Ripple factor | ≤5% | ||
| Test time | 0.0s~999s 0= continue | ||
Applying Illustration
Preparation of test
Avoid electric shock
Please wear insulated gloves and stand on the insulation pad when operate the hipot tester. Note: the withstanding voltage of insulating gloves and insulated pad should be twice than the maximum output voltage of the tester.
Set the alarm value of leak current and test time
Please refer to the seventh chapter.
Connect the DUT(Device Under Test) to the tester
Please confirm the following before connection:
1, The tester is shut down or on the reset status;
2, The high voltage indicator is not light;
3, The voltage window show 0;
Connect the current measurement loop terminal firstly then connect DUT to the high voltage terminal.
Test mode
Manual test
If the test time is set to 0, the tester will not judge the test time during testing. The tester is in a continuous test condition in the test.
After the "START" button is pressed, the test light will be lit; adjust the output voltage knob at a desired value, the tester will do the test continuously. During the test, if the test current is great than the preset value, the tester will alarm with sound and light. Press the "STOP" key, the alarm will be stopped. Replace the DUT and continue to test. After testing, press the “STOP" key to stop the test.
Automatic test
If the test time is not set to 0, the tester will enter into the automatic test. After the "START" button is pressed, the test light will be lit; adjust the output voltage knob at a desired value, the tester will do the test. During the test, if the test current is great than the preset value, the tester will alarm with sound and light. Press the "STOP" key, the alarm will be stopped. If the test time is return to 0 and the tester is not alarm, the test is passed and the DUT (Device Under Test)T is eligible.
If you are satisfied with this instrument, please tell your colleagues; if you are dissatisfied with this instrument, please call (0312) 6775656 and tell us-Baoding Huazheng Electric Manufacturing Co., Ltd., our company will definitely satisfy you!
Before the operation, please read carefully and understand the instructions described in the content. After read the instructions on the vicinity of the operator when needed to read. When the tester is moved from one workplace to another, please send user manual together with the equipment, so as not to lose.
* If you find that missing page manual or instructions to be contaminated, please contact with the dealer of Allwin company immediately to purchase.
* With improved equipment capabilities, software upgrades, instruction manual will also continue to be improved and upgraded. Note that the test instruments and manual versions of the software.
Accessories & Maintenance
| No. | Item | Qty |
| 1 | Power cord | 1 |
| 2 | Test clip | 1 |
| 3 | User manual | 1 |
| 4 | Calibration report | 1 |
| 5 | Qualified certificate | 1 |