HZ1141 Transformer Oil Gas Content Tester
I.Overview
In order to prevent transformer failure better and more comprehensively, the power industry has set higher requirements for the safe operation of transformers at all levels, and the test requirements for the gas content of transformer oil have also increased accordingly.
The insulating oil gas content tester is based on the DL423-91 power industry standard and relevant national standards. It uses imported high-precision sensors and the latest sensor technology. The determination of gas content in ultra-high pressure oil-filled equipment is an ideal test equipment in the laboratory.
II.Main Tech Data
1. High-precision electronic differential pressure sensor: the resolution shall be no less than 0.1%。
2. Minimum detection limit: no higher than 0.2%。
3. Accuracy: the relative error of the results of repeated measurement shall not exceed the following values:
| Gas content in oil (volume fraction),% | relative error,% |
| < 0.5 | 10 |
| 0.5 – 1.0 | 8 |
| 1.0 – 3.0 | 5 |
| > 3.0 | 3 |
4. Temperature control range of degassing chamber: from 5℃ to 70℃ above room temperature
Instrument Accessories List
No. | Name | Qty |
1 | Gas content measuring instrument | 1 set |
2 | user's manual | 1 copy |
3 | power cable | 1 pc |
4 | Vacuum pump and connector | 1 set |
5 | 3mm stainless steel tube | 1.5 meters (connected with gas content tester and vacuum pump 0.5 meters (connecting the oil drain port to the oil drain container) |
6 | Syringe connector | 1 set(Connect the oil inlet to the sampling syringe) |
7 | 10A fuse | 3 pcs |
10 | M8 × 1 nut | 10 pcs |
11 | Seal | 10 pcs |
12 | Syringe holder | 1 pc(installed on the left rear pillar of the gas content tester) |
13 | 1mL calibration syringe | 1 pc(For calibration equipment) |
Supporting Equipment Required for this Instrument
1. A rotary vane vacuum pump (distributed randomly).
2. Air source (built-in)